- Throughput: Up to 100 samples/hour
- Reagents: 4 required reagents; environmentally friendly methanol- and ethlylene-glycol-free
- Sample Volumes: 165 μL aspiration volume; 0.5 mL dead volume. Pre-dilute with x5 dilution mode; 50 μL whole blood into 200 μL of Diluent
- Power: Dual identical power computers if in a workcell.
DxH 900 SPM: 520 W (1775 BTU/hour), Monitor: 35 W (120 BTU/hour)
DxH Server Computer: 160 W (546 BTU/hour) - Weight: 254kg (560 lbs)
- Dimensions: W x H x D: 75.57 x 174 (cover closed)/194 (cover lifted) x 82.20 (including removable back panel)/79.25 (excluding removable back panel)
- UPS: 3 KVA